Nonlinear Spectroscopies
Nonlinear spectroscopy results when a light field perturbs the optical properties of a molecule so that the subsequent light fields see changes in the molecular states. It is important that the different light fields interact with the molecular states over time periods that are shorter than periods associated with dephasing and population relaxation so that the effects of the first interaction are not lost. By exciting a state with one light field and probing it at a later time, nonlinear experiments allow one to investigate how quantum states evolve in time and how different states are related to each other. We must first understand the evolution of coherences and populations that characterize the different nonlinear methods. We use second order spectroscopies as an example.
Second
Harmonic
Generation (SHG), Sum Frequency Generation (SFG), Difference Frequency
Generation (DFG), and Optical Rectification
[back
to the top]
The second
order nonlinear spectroscopies of SHG, SFG, DFG, and optical
rectification are
the simplest examples of nonlinear molecular spectroscopies and will
serve as
our example of diagrams that describe nonlinear spectroscopy. They are three wave mixing (TWM) methods. Second
order spectroscopies vanish for isotropic samples and are therefore
surface
selective. Figure 1a shows
the
flow of coherences that describes all
the
processes. The second order nonlinear spectroscopies involve two
transitions
from the initial ground state population, gg,
to cg for SFG, ba
for DFG, or back to gg
for optical rectification. SFG is a parametric process where the final
emitting
coherence involves the initial state, g,
so the molecule returns to the ground state after the cg coherence
emits. It
has only a single coherence pathway, gg→bg→cg.
SHG is a special case of SFG where the two excitation frequencies are
identical. DFG is a nonparametric process where the final emitting
coherence, ba, does not involve the
ground state so
the molecule is left in an aa
population after the ba coherence
emits. It has two coherence pathways that interfere, gg→bg→ba
and gg→ga→ba.
Optical rectification is a parametric process with one pathway, gg→bg→gg. If the
excitation beam
frequencies are labeled ω1
and ω2,
SHG, SFG, DFG, and
optical rectification have output frequencies at 2ω1
(or 2ω2),
ω1+
ω2,
ω1-
ω2,
and dc, respectively.
SFG and DFG
spectra result from scanning one of the excitation frequencies while
monitoring
the intensity of the output signal from the final coherence. The final
coherence emits light at a different frequency than the excitation
beams so it
is simple to spectrally discriminate between the excitation beams and
the
output signal. The output intensity increases when the excitation
frequencies
match molecular resonances with vibrational and/or electronic states.
Figure 1b
shows the WMEL diagrams that describe the possible resonances for the
three
processes. Figure 1c shows Feynman diagrams that contain the same
information
more explicitly. Here, the two vertical lines represent the ket (left)
and bra
(right) states as time advances vertically from the bottom.
Interactions are
represented by sloped lines. The quantum states before and after an
interaction
are indicated by letters. Lines sloped downward from a ket or bra
vertex are
absorption events (
for
ket side interactions
and
for
bra side
interactions) and lines sloped upward from a vertex are emission events
(
and
for
ket and bra side
interactions, respectively.) The resonance enhancements occur when the
arrows
or combination of arrows match the frequencies of the coherences. The
coherences are defined by identifying the ket and bra states at a given
time
(eg. after the second interaction, the ket-bra
states are cg for SFG
and ba for DFG in figure 1c).
Resonant
enhancements occur when a combination of excitation frequencies matches
the
coherence frequency.
The
relative intensity of the different TWM methods depends on the phase
matching
conditions. The polarization induced in any given molecule will have
all the
possible frequency components formed by linear combinations of the
excitation
frequencies but the observed emission at these frequencies will depend
on how
the phases of the emission from the polarizations of all the molecules
add
together. Phase matching requires
,
,
,
and for
SHG, SFG, DFG, and optical rectification, respectively.
On a surface, the perpendicular component of
is
unimportant and phase matching requires matching the
component parallel to the surface.
Coherent
Raman
Spectroscopies- coherent anti-Stokes Raman (CARS), stimulated Raman
(SRS), impulsive
stimulated Raman (ISRS), inverse Raman (IRS), Raman gain and loss,
coherent
Stokes Raman (CSRS), multiply enhanced nonparametric (MENS), multiply
enhanced
parametric (MEPS) [back
to the top]
The coherent
Raman spectroscopies are four wave mixing processes where three
excitation
fields create the output coherence and one of the intermediate
coherences
involves a vibrational state. The WMEL diagrams for the different
methods
are
sketched in Figure 2. States a, b, and
c are typically vibrational,
electronic, and a vibrational state of the electronic state,
respectively.
Often in Raman methods, the electronic states are so energetic that the
excitation fields are far from resonance. In this case, the states are
indicated by dotted lines and are called virtual states, even though
they are
actually simply off-resonant molecular states. If the excitation fields
are
resonant, the methods are now called resonance-Raman methods. The
output
intensities are typically 103-104
larger because of the
additional resonances.
Both the coherent anti-Stokes
Raman spectroscopy
(CARS) and coherent Stokes Raman
spectroscopy (CSRS,
pronounced
“scissors”) typically use two excitation beams
labeled ωL
and ωS
(indicating laser field at ωo
and Stokes field at ωo–ωv) to
create an output at
the anti-Stokes frequency relative to ωL,
2ωL-ωS or
the
Stokes frequency relative to ωS, 2ωS-ωL,
respectively (see figure 2a,b). Phase matching is important
for optimizing
the output intensity. In CARS, the phase matching is
and
in CSRS, the phase
matching is
.
CARS is used almost exclusively for coherent Raman
experiments because its output is at higher frequencies than the
excitation
sources and is therefore less sensitive to fluorescence that might be
created
by the excitation beams.
Experiments
are performed by changing one excitation frequency relative to the
other and
monitoring the output intensity. The output intensity increases by ~103
when ωL-ωS
is
resonant with a strong vibrational transition. With monochromatic
excitation
sources, the frequency is changed by scanning the frequency of one
relative to
another. Multiplex CARS is performed by using a broadband source for
one
excitation frequency and
spectrally
resolving the output frequencies, typically with a monochromator and a
CCD
camera. This allows one to acquire complete vibrational spectra on a
single
laser shot. For ultrafast multiplex CARS spectroscopy, a chirped pulse
(pulses
where the frequency changes rapidly during the pulse duration) with a
several
picosecond duration and a fast femtosecond pulse (~100 fs) are used for
ωL
and ωS,
respectively. The fast ωS
pulse overlaps in time with only part of the longer ωL
pulse and selects the
frequencies present in the ωL
pulse at the time of overlap. The broad band of frequencies in the ωS pulse
can achieve
resonance with many vibrational states at ωL-ωS so a series of ag vibrational coherences are created
(see figure 2a) whose FID
occurs at specific vibrational frequencies. The last excitation is also
the
chirped ωL
pulse so the
vibrational coherences will interact with the frequencies present in
the
chirped pulse during the time they overlap and create output
frequencies at
.
Detection of the signal with a monochromator and a CCD
camera then resolves the anti-Stokes frequencies from the different
vibrational
coherences.
CARS is attractive for spectroscopy because the output signals are large and directional, so interference from incoherent fluorescence can be discriminated against both spectrally using a monochromator and spatially by using an aperture to define the CARS output beam. Phase matching is usually important to achieve reasonable signal levels.
One of CARS
most important limitations is the presence of signals that arise from
nonresonant electronic states since state a
in figure 2a can be a resonant vibrational state or a nonresonant
electronic
state. Electrons are much lighter than nuclei so the polarization
induced by an
electric field is usually dominated by the electronic component of the
polarization rather than the nuclear component. Resonance with a
vibrational
state raises its contribution to the polarization so it can dominate
over a
nonresonant electronic polarization but the nonresonant electronic
polarization
must always be considered. The relative contributions are described by
the nonlinear
susceptibilities,
.
In the steady state, the vibrational contribution is
(3)
and its importance depends on the
resonant enhancement from
the detuning factor
.
In CARS, a strong Raman transition has a
peak that is ~25x
larger
than
.
A number of methods have been developed to discriminate against the nonresonant background. Since the nonresonant electronic polarization decays almost instantaneously after the excitation field is turned off, ultrafast pulses can be used to excite a vibrational coherence. Since vibrational coherences typically live for ~1-10 ps, delaying the third pulse relative to the first two can strongly discriminate against any nonresonant electronic coherence but still excite the vibrational coherences to create the output coherence.
Polarization
techniques can also be used for discrimination. The output CARS signal
is
polarized if the excitation beams are polarized. A polarizer can be
adjusted to
block the signal beam from being detected if the exciting frequencies
are not
resonant with a vibrational state. However, if the excitation
frequencies are
changed to a vibrational resonance, the output polarization can change
because
the
and
tensors
are different,
their transition moments have different dependences on the polarization
of the
exciting electromagnetic fields. A portion of the output signal can now
pass
through the polarizer and be detected.
Figure 2c shows their WMEL diagram for Raman gain or stimulated Raman and Raman loss or inverse Raman spectroscopies. The most important difference between these processes is that the output frequency matches an excitation frequency, so the output beam at ωS can interfere (heterodyne) with the excitation beam at ωS. The transitions involving the ωL beams are absorption transitions, so the intensity of the transmitted beams decreases, while the transitions involving the ωs beams are stimulated emission transitions, so the intensity of the transmitted beams increases. There is no need for phase matching considerations when the output frequency matches one of the excitation frequencies since the phases have to be identical. Consequently, these spectroscopies use collinear beams.
Raman gain and loss spectroscopies are performed by changing one excitation beam frequency while monitoring the increase in the transmitted ωS intensity or decrease in the transmitted ωL intensity, respectively. If there is only a single excitation frequency, ωL present and its intensity is high, vacuum fluctuations in the quantized electromagnetic field can act as an ωS beam to stimulate emission at ωS so an ωS beam can be created from the vacuum. The creation of a beam at ωS is stimulated Raman scattering (SRS). Usually, the Raman transition with the largest transition moment dominates the process. Inverse Raman spectroscopy (IRS) is performed with a monochromatic excitation frequency, ωS, and a broad band source at ωL. The transmitted light at ωL is measured with a monochromator and multiplexing detector like a CCD and absorption lines appear in the dispersed output at ωL from vibrational transitions at ωL-ωS. One way to view these processes is to realize that in the absence of a sample, the fields at ωL and ωS would be independent. However, if a sample is present, the fields can exchange energy using the sample as an intermediary. If the ωS beam is initially absent, the ωS beam intensity grows and the ωL beam intensity decreases.
An ultrafast pulse can have a sufficiently wide range of frequencies (a 100 fs pulse has a frequency bandwidth of ~350 cm-1) that it alone can provide both the ωL and ωS frequencies and therefore excite vibrational states by stimulated Raman scattering. This process is called impulsive stimulated Raman scattering (ISRS) and it represents an important way to excite many vibrational states simultaneously.
Stimulated
Fluorescence,
Pump-Probe Spectroscopies, and Time Resolved Pump-Probe Spectroscopies
[back
to the top]
Fluorescence and resonance Raman spectroscopies are closely related. Both are FWM processes and both involve similar excitation frequencies. They differ in the time ordering of the exciting fields and they differ in whether the intermediate state is a coherence or population. Figure 2d shows an example WMEL diagram for stimulated emission of fluorescence. The initial ket and bra side excitations create an excited bb population. This population is stimulated to emit by interaction with the third field which forms the ba output coherence. In contrast, the intermediate state in stimulated Raman spectroscopy is an ag coherence.
Pump-probe methods are
also closely
related. These methods almost always use FWM processes. Example WMEL
diagrams
of single color pump-probe methods are shown in Figure 3. Pump-probe
methods
involve two overlapped excitation
pulses-
an initial pump pulse that
creates
the first two interactions in each diagram in figure 3 and a probe
pulse that
creates the third interaction leading to the final output coherence.
Since the
output field has the same frequency as the probe field, the two fields
interfere. The transmitted intensity of the probe field can either
increase
because the pump induced stimulated emission (figure 3a) or ground
state
bleaching (figure 3b) or decrease because the pump created an excited
state
population that allow excited state absorption. The net change depends
on the
relative strength of all these processes.
Two color pump-probe experiments use different frequencies for the pump and probe pulses. Figure 3d-f shows the two color pump-probe pathways. The first two electric field interactions represent the pump and create a population. The third interaction is the probe that creates the output coherence. The bleaching, stimulated emission, and excited state absorption pathways are shown in figures 3d; 3e; and 3f, respectively.
The quantum states involved in the pump or probe steps can be electronic/vibronic or vibrational. Electronic states are excited through uv/visible absorption while vibrational states are excited by infrared or Raman (or stimulated Raman) transitions. Consequently, time resolved (TR) pump-probe methods are classified as TR-UV-UV, UV-IR, IR-IR, UV-Raman and IR-Raman, etc. depending upon whether an electronic or vibrational state is pumped or probed. Since Raman is actually a four wave mixing process, the UV-Raman and IR-Raman pump-probe experiments are actually six wave mixing processes. Note that if the two pump interactions are created instead by separate controllable beams, these pump-probe pathways become rephasing stimulated photon echo (figures 4c, e, g) or nonrephasing (figures 4d, f, h) pathways.
Photon Echo, Stimulated Photon Echo, Transient Grating, Reverse Photon Echo, and Reverse Transient Grating Spectroscopies [back to the top]
Photon echo was one the
first examples
of an optical analogue of NMR. Figure 4a,b show the two photon echo
WMEL
diagrams. An initial pulse creates a ga
coherence after the first interaction. After a delay time, τ, a second pulse induces the
second and third interactions to
create an ag coherence that
reemits.
Phase matching is important. For photon echo, the phase matching
condition is
where
the subscripts
indicate the time ordering of the two pulses. The usefulness of photon
echo
rests on having spectral transitions that are inhomogeneously broadened
so
molecules that have different environments or conformations have a
distribution
of different transition frequencies. The temporal dependence of the
excited ga coherences is
which
becomes
after
the τ
delay. The temporal dependence of the
final ag coherence is
,
conjugate to the ga
coherence. After a delay of τ,
the
net phase will be zero, regardless of the ωag
frequency.
At this point in time, all of the coherences will
be again in phase and the re-emission becomes completely coherent (it
scales as
N2). The
large output
intensity that results at time τ
is
called the echo. By measuring the echo intensity as function of delay
time, one
can measure the dephasing rate (Γag)
of the ga coherence.
Stimulated photon echo
experiments are
three pulse experiments so the three excitation interactions in figure
4
occur at
different times. The phase matching condition is
.
Figure 5a-c shows there are three pathways- a) ground state
bleaching, b) stimulated emission, c) excited state absorption. The
sign of the
polarization is (-1)n where n
is the
number of bra-side interactions so excited state absorption has the
opposite
sign from the other pathways, that is it creates intensity decreases
rather
than increases. After the second interaction in figure 4a,b, one has
either a gg ground state population
(figure 4a) or
an aa excited state population
(figure 4b). After a delay time T, the third pulse
arrives
to create the ag output coherence.
Measuring the
stimulated photon echo intensity as a function of the T
delay time measures the population changes of the ground and
excited state populations.
Figure 4e,f
shows the WMEL diagrams for transient
grating experiments. Transient grating methods use two pulses
that occur at
the same time but are angled relative to each other. The phase matching
condition is
where
the subscripts
again indicate the time ordering. These pulses are represented by the
first two
interactions in fig. 4e,f. They create gg
(figure 4e) and aa (figure 4f)
populations that are spatially modulated (
)
to form a grating. The third pulse acts as a probe to
create the ag output coherence. Its
direction corresponds to its reflection off the grating created by the
first
two pulses. By measuring the transient grating intensity as a function
of the
delay between the two excitation pulses, one measures the population
decay of
the ground and excited state populations. A second delay can be
introduced by
separating the first two pulses in time. The transient grating
experiment now
looks very similar to the stimulated photon echo experiment. The most
important
difference lies in how they respond to inhomogeneous broadening. Note
that the
coherence formed after the first pulse is an agcoherence
and so is the coherence formed after the last pulse. The phase
difference created during the first delay no longer cancels the phase
change formed by the last coherence; they now add, and
rephasing
never occurs. This form of transient grating experiment is called a
non-rephasing pathway.
Methods
called reverse photon echo and reverse transient grating also exist.
These methods are based on reversing the time orderings of the two
pulses. In
photon echo, the last two interactions occurred simultaneously. In
reverse
photon echo, the two simultaneous interactions now occur first. The
phase
matching is
.
Similarly, the first two interactions were simultaneous in
transient grating methods. In reverse transient grating experiments,
the two
simultaneous interactions now occur last. The phase matching is
.
Again, the subscripts indicate the time ordering. Typical
WMEL for these methods are shown in figure 4d and h. These methods have
not
been used appreciably except in the developing field of coherent
multidimensional spectroscopy.
Incoherent
Fluorescence
and Raman Spectroscopies and Their Relationship to Coherent
Spectroscopies [back
to the top]
Fluorescence and Raman
(affectionately known as COORS or
Common Old Ordinary Raman
Spectroscopy) spectroscopies are also four wave mixing nonlinear
spectroscopies
but they are incoherent, i.e. there is no phase relationships between
excited
molecules in different parts of a sample. The WMEL diagrams for Raman
and
fluorescence spectroscopy are still represented by figures 2c and d,
respectively. However, instead of the ωS
excitation being caused by a real excitation beam, it is instead caused
by the
vacuum fluctuations which have no long range phase relationships so the
coherences they stimulate are incoherent and therefore not directional.
The
emission intensity then scales as the concentration N.
Coherent
Multidimensional Spectroscopies
[back
to the top]
Ever since Feynman et al showed that the transitions between two quantum states were equivalent to the spin excitations in NMR experiments, there has been great interest in developing the optical analogues of NMR, particularly because of the power of NMR to probe complex systems with great selectivity. The invention of the laser stimulated a great deal of work to develop the optical analogues because the laser provided the required coherent source.
Early experiments closely followed NMR methods. π/2 and π pulses were delivered to samples by carefully controlling the phase and intensity of the excitation. It was difficult to preserve the phase information of the coherences because the vibrational and electronic coherences dephase ~9 orders of magnitude more quickly than NMR coherences because of the stronger interactions with the environmental thermal bath. These dephasing rates are much larger than typical Rabi frequencies, so it was difficult to create coherences with π/2 and π pulses owing to the time required being longer than that permitted by the dephasing times. Success was achieved only by making the pulses stronger so the Rabi periods were shorter or increasing the dephasing times so the dephasing times were comparable to the Rabi period. Typically, the dephasing times were increased by cooling the sample to extremely low temperatures. Although these approaches allowed limited studies, they did not provide a viable approach for measuring typical samples.
Coherent
multidimensional spectroscopy (CMDS) has emerged as a technique with
widespread
use because it was di
scovered that phase matching and time ordering of
the
excitation pulses could isolate particular coherence pathways just as
well as
supplying a series of phased π/2 and π pulses. For
example, a molecule in its
ground state can interact with a field to create a coherence either
through
absorption on its ket- or bra-side. However, if the experiment uses a
stimulated photon echo phase matching geometry
,
the first interaction must be a bra-side absorption; so for
a two state system, the pathways are restricted to the two shown in
figure 5.
CMDS can be
based on time domain or frequency domain methods. In time domain
methods, one
directly measures the temporal oscillations of the coherences and then
performs
a Fourier transform to obtain a frequency spectrum. If excitation
pulses were
available that were very short compared with the period of a
coherence’s
oscillation, the measurement could be performed by scanning the delay
of the
excitation pulse relative to the coherence being measured and one would
immediately obtain the temporal phase oscillations of the coherence.
Unfortunately, the excitation pulses are not that short so one must use
a local
oscillator. A local oscillator usually is a fourth field that is almost
identical to the signal beam. The electric fields of the local
oscillator and
the signal heterodyne so the intensity depends on
.
If the time delay between the local oscillator and the
signal is changed, one directly measures the oscillations of their
relative
phases. A Fourier transform will provide the frequency domain
spectrum. If
the delay between the first two excitation beams changes, the signal
beam will
reflect the changes in phase of the coherence created by the first beam
and
these in turn will be measured by the local oscillator as well. Fourier
transform of these variations provide a second frequency axis for the
2D-IR
spectroscopy. If the first two excitation beams create a coherence, the
delay
between the second and third excitation beams will scan the temporal
phase
changes of that intermediate coherence and Fourier transformation will
provide
a third frequency dimension. It is very important that the phase
relationships
between the excitation beams remains stable over the entire measurement
time.
This factor requires that the multiple excitation beams be derived from
a
single source that acts to define the phase of each beam. It also
requires that
the relative path lengths within the experimental system are
interferometrically stable to a small fraction of a wavelength.
Finally, in a
pure time domain experiment,
all the
quantum states are excited impulsively and one measures all of the
coherences
simultaneously through the temporal oscillations. Practically, the
range of
quantum states that can be measured depends on the excitation pulse
bandwidth.
Typically, the pulse widths are ~50-100 fs and these bandwidths excite
quantum
states over a range of ~150-300 cm-1.
In frequency domain methods, one scans the frequency of the excitation beams while monitoring the intensity of the output coherence. The intensity is enhanced by each resonance and the enhancements are multiplicative. Phase coherence is again required but it is only required during the pulse sequence. Long term phase coherence is not required. Finally, in a pure frequency domain experiment, only specific quantum states are excited at any one time so individual coherence pathways and specific states are selected. The range is limited only by the tunability of the excitation sources. There are advantages to working in a mixed frequency/time domain where the excitation pulses are long enough to excite individual quantum states and avoid the difficulties involved with maintaining long term phase coherence but they are also short enough that one can measure the temporal dynamics of the coherences and populations of the quantum states.