Implementing Photodissociation in an Orbitrap Mass Spectrometer

TitleImplementing Photodissociation in an Orbitrap Mass Spectrometer
Publication TypeJournal Article
Year of Publication2011
AuthorsVasicek, LA, Ledvina, AR, Shaw, J, Griep-Raming, J, Westphall, MS, Coon, JJ, Brodbelt, JS
JournalJournal of the American Society for Mass Spectrometry
Volume22
Pagination1105-1108
Date PublishedJun
Accession NumberISI:000290451600018
KeywordsBiochemical Research Methods, Chemistry,, Chemistry, Analytical, electrospray-ionization, HCD cell, infrared multiphoton dissociation, orbitrap, peptide, photodissociation, Physical, protein, proteomics, quadrupole ion-trap, spectroscopy
Abstract

We modified a dual pressure linear ion trap Orbitrap to permit infrared multiphoton dissociation (IRMPD) in the higher energy collisional dissociation (HCD) cell for high resolution analysis. A number of parameters, including the pressures of the C-trap and HCD cell, the radio frequency (rf) amplitude applied to the C-trap, and the HCD DC offset, were evaluated to optimize IRMPD efficiency and maintain a high signal-to-noise ratio. IRMPD was utilized for characterization of phosphopeptides, supercharged peptides, and N-terminal modified peptides, as well as for top-down protein analysis. The high resolution and high mass accuracy capabilities of the Orbitrap analyzer facilitated confident assignment of product ions arising from IRMPD.

Short TitleJ. Am. Soc. Mass Spectrom.